Resumen
It is not always possible to accurately describe the relationship between a set of calibration points with a rectilinear
curve, even by decreasing the working range. Instead of the linear regression analysis, a least-squares fit to a
second-order polynomial is applied (see test for linearity in 4.1.3 of ISO 8466-1:1990[1]). Using this fit, it is possible
to calculate not only the calibration function but also the confidence interval associated with it.
This part of ISO 8466 is intended primarily for use in method development and may not necessarily be applicable to
all routine analyses.
Informaciones generales
-
Estado: PublicadoFecha de publicación: 2001-11Etapa: Norma Internacional confirmada [90.93]
-
Edición: 2Número de páginas: 12
-
Comité Técnico :ISO/TC 147/SC 2ICS :13.060.01
- RSS actualizaciones
Ciclo de vida
-
Anteriormente
RetiradaISO 8466-2:1993
-
Ahora